DCU7 Full Digital Color Doppler Ultrasound scanner
1. System Application
DCU2 Full Digital Color Doppler Ultrasound Scanner is designed for comprehensive diagnostic capabilities, particularly in the abdomen (including GYN and OB), cardiology, vascular, and small parts. This versatile device offers various scanning modes including 2D, CFM, PDI, PW, and M, ensuring the highest resolution and sensitivity.
Supports convex probes, linear probes, and transvaginal probes, among others.
2. General Specification
2.1 Dimensions
Main unit dimensions: 660mm(H) x 500mm(W) x 1200-1400mm(H) (adjustable LCD holder height)
Net Weight: 35 Kg (excluding probe)
2.2 Power Supply
Dual power-supply system:AC110/230V -15%~+10%(self-adaptive supply voltage) 60/50Hz ±1Hz
2.3 Display
15" high-definition & flicker-free color medical LCD monitor
3. System Description
3.1 Probe Specifications
Probe specification |
3.5MHz multi-frequency convex probe |
7.5MHz multi-frequency linear probe |
6.5MHz multi-frequency transvaginal probe |
Scanning angle |
60° |
/ |
135° |
elements |
80 |
80 |
80 |
frequency scope |
2.0MHz-5.0MHz |
6.0MHz-9.0MHz |
4.5MHz-9.0MHz |
4 multi-frequency |
2.0/2.5/3.5/5.0 |
6.0/6.5/7.5/9.0 |
4.5/5.5/6.5/7.5 |
Max scanning depth |
≥190mm |
≥60mm |
≥60mm |
Blind zone |
≤5mm |
≤3mm |
≤4mm |
3.2 Operating Modes
B, B/B,4B mode
M,B/M mode
CFM
PDI
PW
THI
4. Measurement / Calculation
4.1 General Measurements
B Mode Measurements: distance, area (distance, ellipse, trace methods), circumference (distance, ellipse, trace methods), volume (two-axial method), angle, histogram, sectional view, stenosis rate (length and area), and depth.
M Mode Measurements: depth, slope, heart rate, circumference
D Mode Measurements: velocity, heart rate, flow velocity, time acceleration, Pulse/ Resistance Index, maximum differential pressure, PG mean, ratio, stroke volume, flow rate.
4.2 Gynecology Measurements and Analysis :uterus, endometrium, ovary, cervix, follicle.
4.3 Obstetric Measurements and Analysis
BPD, CRL, GS, FL, HL, TAD, TTD, APTD, HC, AC, EDC, FT, LV, THD, TCD, OFD, GA, EFW, LMP, AFI, and calculates GA, EFW, EDD, growth curve, with various country-specific formulas.
4.4 Urology Measurements and AnalysisProstate volume, Residual Urine Volume.
4.5 Cardiology Measurements and Analysis
4.6 Measurement Reports
Obstetric measurement report,
gynecology measurement report,
cardiology measurement report,
urology measurement report, and other measurement reports.
Automatically stores measurement results and generates reports.rt
Standard configuration: 1. High-resolution color display 2. Advanced imaging algorithms 3. Ergonomic design for ease of use 4. Multi-frequency transducers 5. Comprehensive measurement and analysis tools 6. Digital storage and transfer capabilities
Main unit:1 pc |
Double probe sockets, automatically identify probe; |
3.5MHz multi-frequency abdomen convex probe:1pc |
7.5MHz multi-frequency high linear probe:1pc |
CD: 1pc |
Reticle: 1pc |
Optional configuration: 1. 3D/4D imaging technology 2. Portable trolley with integrated storage 3. Enhanced Doppler sensitivity modules 4. Wireless connectivity for seamless data sharing 5. Extended warranty and service plans
Video Printer(sony 897-MD) |
6.5MHz multi-frequency trans-vaginal prob |